Class III Laboratory Biological Safety Biosafety Cabinet Class III Bsc Series

General Details

Brand BIOBASE
Product origin CHINA
Supply capacity Factory Direct Supply
Advantage:

  • With memory function in case of power-failure.
  • UV lamp ensures work area sterilization effectively.
  • An angled cabinet front ensure an ergonomic working posture.
  • Exhaust air is double-filtered through high-quality ULPA filters with typical efficiency of 99.9995% for 0.12μmparticles.

Class III Biosafety Cabinet is totally enclosed and gas-tight, it also can meet the requirement of operating Class I,II,III,IV pathogenic factor. It can be used in P3,P4 laboratory.

Class III Biosafety Cabinet is totally enclosed and gas-tight with ULPA filtered supply and exhaust air. Work is performed with long-sleeved gloves. The cabinet is kept under negative pressure of at least 120Pa, and airflow is maintained by a dedicated exterior exhaust system. It can protect the operator, product and environment.

Model
Model BSC-1100IIIX BSC-1500IIIX
External Size(W*D*H) 1540*880*2000 mm 1790*880*2080 mm
Internal Size(W*D*H) 910*650*660 mm/l 1165*650*660 mm
Pass Box Size(W*D*H) Internal Size:430*330*355mm
External Size:575*425*495mm
Gloves One pair of 800mm butyl rubber gloves
Front Window 8 mm toughened glass, anti-ultraviolet radiation
Display LCD display
Pre-filter Polyester fiber, washable
Pressure ≥negative120 Pa
ULPA Filter Air supply filter: Efficiency 99.9995% at 0.12μm
First exhaust filter: Efficiency 99.9995% at 0.12μm
Second exhaust filter: Efficiency 99.9995% at 0.12μm
Noise ≤67dB
Alarm Audio and visual alarm, abnormal pressure; filter replacement
Work Zone Material 304 stainless steel
Illumination ≥ 1000 lux
Consumption 1200W 1300W
LED Lamp 12W*2 16W*2
UV Lamp 18W*2 ,8W*1 30W*2, 8W*1
Power Supply AC220V±10% ,50/60 Hz;110V±10%,60H
Gross Weight 350kg 450kg